Normal flow vs. murmur

The term "cardiac murmur" evokes an image of the heart whispering softly, and this is not far from the truth. Cardiac murmurs are extra heart sounds that can give clues to certain disorders. This figure lets you listen to one type of murmur and see the type of blood flow that makes it. Press Play to start the heart and hear it; headphones help, and the Sound button toggles the sound. Choose aortic stenosis to hear the murmur of a narrowed aortic valve.

Some key things to keep in mind when looking and listening:

Did you know? The highest grade of heart murmur has historically been called a "thrill", a term used when there is enough turbulence that the murmur can be felt as a vibration with a hand on the chest wall. Such murmurs are rare, particularly today, but there are stories of medical students and doctors getting in trouble for using the term in front of patients: "Come over here, this is quite a thrill!" You can imagine.

Questions this figure helps answer

The figure explains the physiology behind these questions. It cannot tell what causes any one person's murmur or whether it matters; a clinician determines that.

The model behind this figure

The figure comes from the model of the heart and circulation behind the pressure-volume figures, with the aortic valve as an opening of a given area. Blood passing the opening must speed up, and that costs a pressure drop of about 4v² mmHg at a speed of v meters per second, the simplified Bernoulli relation used in echocardiography. A normal valve opens to 4 cm², and the narrowed valve to 1.2 cm², moderate aortic stenosis. With the same stroke volume, blood leaves the narrowed valve at up to 3.7 m/s instead of 1.8, and the ventricle's peak pressure rises from 123 to 155 mmHg to push it through.

The murmur comes from the jet. Above about 1.8 m/s, blood leaving the valve breaks into eddies as it enters the wider aorta. The murmur's loudness rises with the jet's top speed, and over the beat it follows the flow through the valve. The heart sounds are generated from the model: the first and second heart sounds as short low thumps when the valves close, and the murmur as low rushing noise. Their timing, shape, and pitch were matched to recorded heart sounds from the University of Michigan Heart Sound and Murmur Library and from Wikimedia Commons.

Normal and aortic stenosis beat 60 times a minute, as in the original figure, which leaves room to hear each sound. High flow beats 90 times a minute with a stroke volume of 90 mL instead of 82, so 8.1 liters of blood a minute pass a normal valve instead of 4.9, as the heart works harder in fever, anemia, pregnancy, or exercise. The phonocardiogram always shows one whole beat, so at 90 a minute it covers a shorter time. The drawing is the HeartPhys figure, redrawn, and as in the original the ventricles contract and fill with each beat.

The figure explains physiology and is not clinical advice. It shows the left side of the heart; it leaves out murmurs from the right side and the splitting of the second heart sound with breathing, and it shows aortic stenosis at one size, without the thickened ventricle of long-standing disease.

Using this figure with an AI assistant

If you use an AI assistant that can see this page, whether it is built into your browser or opens the page in its own browser, you can ask it about this figure, ask it to show you what happens in a particular case, or ask it to take you through one of the tours. It can work the figure itself, point out its parts, and explain what you see. It cannot hear the sound. The rest of this section is for the assistant.

An AI assistant can use this figure in three ways. In a browser with WebMCP, the page offers its actions as tools. In the page's scripts, the same actions are in window.physiologyFigures, with a description of the controls and what they mean, the values shown, the model's limits, and the Show Me tours. Both do what the page's buttons and slider do, wait until the figure shows the result, and return the exact values, and they can also point out parts of the figure. An assistant can also click the buttons and drag the slider, but then it has to read the results off the screen. The page address always holds a link to the moment shown, such as ?condition=stenosis&moment=450.